Products

ECD
Electrochemical Detectors
LC/ECD
UHPLC with Electrochemical Detection
EC/MS
Electrochemistry with MS Detection

ROXY Potentiostat

ROXY

ROXY™ Potentiostat.

ROXY™ Potentiostat controls redox reactions

 

  • Stand alone or on-line EC/MS
  • Fully programmable, direct control, scan and activation mode
  • Large voltage range, ± 4.9 V
  • Multiple cell control (standard 2, optional up to 4)
  • Compatible with Reactor™ and µ-PrepCell™

 

With more than 20 years of experience in Electrochemistry (EC), Antec introduces a new, dedicated Potentiostat for on-line EC/LC/MS. The ROXY™ Potentiostat generates metabolites of drugs or xenobiotics, similar to those generated during in vivo metabolic processes, in a significantly shorter time span (seconds vs. days or weeks) without any interfering components (no isolation steps required). The ROXY™ Potentiostat is based on state-of-the-art electronics with a large voltage range of ± 4.9 V, and a push button electrode regeneration program. Operational parameters and external equipment can be controlled through programmable timed events. Consequently, the generation of specific oxidation products, e.g., metabolites, cleavage products, etc., and supreme control of any conceivable Redox reaction is assured.

Two in One!
The ROXY™ Potentiostat is standard supplied with one ReactorCell™, and has the capability of controlling two cells independently. This is ideal for simultaneous use of different working electrodes or potentials (parallel mode) or multiple stage oxidation/reduction of target compounds (serial mode). The ROXY™ Potentiostat can be upgraded to control up to a maximum of four ReactorCells™, and comes fully equipped with a high precision oven for optimal conversion kinetics and reproducibility (ambient up to 45°C).

General specifications

Power

110-240 VAC, 50/60 Hz, 260 VA, autosensing

Operating modes

DC, PAD and Scan

Potential range

between ± 4.90 V in 10 mV increments

Output

between ± 1 V or between ± 10 V (20 bit D/A converter)

Offset

between ± 50% of max. output voltage, 5% steps

Event marker

pulse of 10% of max. output

Auto zero

triggered by keyboard, rear panel TTL input , or RS232C control

RS232C

Full parametric instrument control, data acquisition at 1, 2, 5 and 10 Hz

Injector sensor

Starts system clock at injection

Oven

height 37 cm, from 7°C above ambient to 45°C, accuracy 0.5°C, stability 0.1°C; accommodates column and flow cell(s)

Diagnostics

LCD screen, keyboard and noise (internal dummy cell)

Service mode

system settings & calibration parameters

Config mode

menu for system customisation and optimisation

Firmware

upgradeable via PC (RS232)

Environmental

operating temperature: 4 – 40 deg C, rel. humidity: 20 to 80% non-condensing

Second flow cell

Acquisition and control of 2nd .. 4th flow cell (option)

Regulatory

CE, UL/CSA, RoHS compliant

 

DC mode

Ranges

10 pA – 200 µA in 1, 2, 5 steps

Filter (cut off)

0.5 – 0.01 Hz in 1, 2, 5 steps

Noise

better than 2 pA with a dummy cell (load of 300 MOhm and 0.5 µF) with filter off, Ec +800mV and temperature of 30 oC.

 

PULSE mode

Range

10 nA – 200 µA in 1, 2, 5 steps

Filter (cut off)

0.5 – 0.01 Hz in 1, 2, 5 steps

Pulse times

t1: 100 - 2000 ms; t2: 0 - 2000 ms; t3: 0 - 2000 ms in 10 ms steps

Sample times

20, 40, 60, 80 and 100 ms

SCAN mode

Range

10 nA - 200 mA in 1, 2, 5 steps

Scan rate

1 - 50 mV/s in 1, 2, 5 steps

Cycle

half, full or continuous

 

Events

Events

DC mode (5 files) and pulse mode (4 files), end cycle time, number of cycles and oven temperature. Time-based control of 50 time points as to range, filter, output contacts (2 TTL, 2 relays), auto zero, offset, valve position (if present), and E-cell.

 

Rear panel

I/O connections

Mains, Output, 2 Connectors 15 pins (A, B), manual valve (C), RS232C connector

Physical specifications

Dimensions

44 (D) x 22 (W) x 44 (H) cm = 17.3” (D) x 8.7” (W) x 17.3” (H)

Weight

14 kg without flow cell and column