ROXY™ Potentiostat. |
With more than 20 years of experience in Electrochemistry (EC), Antec introduces a new, dedicated Potentiostat for on-line EC/LC/MS. The ROXY™ Potentiostat generates metabolites of drugs or xenobiotics, similar to those generated during in vivo metabolic processes, in a significantly shorter time span (seconds vs. days or weeks) without any interfering components (no isolation steps required). The ROXY™ Potentiostat is based on state-of-the-art electronics with a large voltage range of ± 4.9 V, and a push button electrode regeneration program. Operational parameters and external equipment can be controlled through programmable timed events. Consequently, the generation of specific oxidation products, e.g., metabolites, cleavage products, etc., and supreme control of any conceivable Redox reaction is assured.
Two in One!
The ROXY™ Potentiostat is standard supplied with one ReactorCell™, and has the capability of controlling two cells independently. This is ideal for simultaneous use of different working electrodes or potentials (parallel mode) or multiple stage oxidation/reduction of target compounds (serial mode). The ROXY™ Potentiostat can be upgraded to control up to a maximum of four ReactorCells™, and comes fully equipped with a high precision oven for optimal conversion kinetics and reproducibility (ambient up to 45°C).
|
Power |
110-240 VAC, 50/60 Hz, 260 VA, autosensing |
|
Operating modes |
DC, PAD and Scan |
|
Potential range |
between ± 4.90 V in 10 mV increments |
|
Output |
between ± 1 V or between ± 10 V (20 bit D/A converter) |
|
Offset |
between ± 50% of max. output voltage, 5% steps |
|
Event marker |
pulse of 10% of max. output |
|
Auto zero |
triggered by keyboard, rear panel TTL input , or RS232C control |
|
RS232C |
Full parametric instrument control, data acquisition at 1, 2, 5 and 10 Hz |
|
Injector sensor |
Starts system clock at injection |
|
Oven |
height 37 cm, from 7°C above ambient to 45°C, accuracy 0.5°C, stability 0.1°C; accommodates column and flow cell(s) |
|
Diagnostics |
LCD screen, keyboard and noise (internal dummy cell) |
|
Service mode |
system settings & calibration parameters |
|
Config mode |
menu for system customisation and optimisation |
|
Firmware |
upgradeable via PC (RS232) |
|
Environmental |
operating temperature: 4 – 40 deg C, rel. humidity: 20 to 80% non-condensing |
|
Second flow cell |
Acquisition and control of 2nd .. 4th flow cell (option) |
|
Regulatory |
CE, UL/CSA, RoHS compliant |
|
Ranges |
10 pA – 200 µA in 1, 2, 5 steps |
|
Filter (cut off) |
0.5 – 0.01 Hz in 1, 2, 5 steps |
|
Noise |
better than 2 pA with a dummy cell (load of 300 MOhm and 0.5 µF) with filter off, Ec +800mV and temperature of 30 oC. |
|
Range |
10 nA – 200 µA in 1, 2, 5 steps |
|
Filter (cut off) |
0.5 – 0.01 Hz in 1, 2, 5 steps |
|
Pulse times |
t1: 100 - 2000 ms; t2: 0 - 2000 ms; t3: 0 - 2000 ms in 10 ms steps |
|
Sample times |
20, 40, 60, 80 and 100 ms |
|
Range |
10 nA - 200 mA in 1, 2, 5 steps |
|
Scan rate |
1 - 50 mV/s in 1, 2, 5 steps |
|
Cycle |
half, full or continuous |
|
Events |
DC mode (5 files) and pulse mode (4 files), end cycle time, number of cycles and oven temperature. Time-based control of 50 time points as to range, filter, output contacts (2 TTL, 2 relays), auto zero, offset, valve position (if present), and E-cell. |
|
I/O connections |
Mains, Output, 2 Connectors 15 pins (A, B), manual valve (C), RS232C connector |
|
Dimensions |
44 (D) x 22 (W) x 44 (H) cm = 17.3” (D) x 8.7” (W) x 17.3” (H) |
|
Weight |
14 kg without flow cell and column |