Dialogue™ is a multi-functional program to control the DECADE II electrochemical detector and ROXY Potentiostat. It is a tool for:
The free version for the SDC is used for control of all operational parameters of the DECADE SDC. There are no tools for further data analysis, therefore Dialogue™ software can not be considered as a fully featured chromatography software package. Dialogue™ software comes as one software package containing functionality for all different versions. Depending on the license dongle PQ/OQ functionality will be unlocked.
| Part no | Description |
| 171.9002 | Dialogue OQ/PQ/ROXY version (for distributors) |
| 171.9005 | Dialogue PQ version |
| 210.9005 | Dialogue for ROXY potentiostat |
| 171.9007 | Dialogue for SDC, free version |
For full trial versions see download section. The trial version is activated by a license dongle to unlock full functionality, no additional software is required.
PQ noise measurement
Dialogue can be used for a Performance Qualification using a dummy cell test. The results is a single page PQ report in Excel. Below a typical stabilization pattern. Using the driftmonitor enables an automatic restart after 3 min in case the noise is out of spec for convenient unattended operation.

Timed events
In ROXY Dialogue the operational parameters of the potentiostat and syringe pump can be controlled through the events table.

Integrated scripting tool for OQ
An OQ script is a series of commands to set parameters and read data from the detector. A script is executed by the script manager in Dialogue™ software and results are printed through Excel in an OQ report. For successful qualification an OQ kit is available containing cables and tools for testing.

Database for audit trial and user rights
An audit trail is available, keeping track of all events initiated by registered users.