Products

ALEXYS HPLC ECDROXY EC MSSystem componentsROXY PotentiostatDECADE II DetectorFlow cellsSoftwareQualification procedures

Autosamplers

AS 110

AS 110 - Perfect fit for high throughput,
routine analysis and small samples.

AS 110 - Fast and robust

 

  • Fast, accurate & reproducible
  • Unique needle concept
  • Sample pick-up down to 1µL
  • Cooled sample tray
  • Well plates or vials

 

The AS 110 is a new generation autosampler. It is a fast, robust and easy to use autosampler without compromising accuracy and reproducibility. The AS 110 includes indispensable features for the highest injection performance. It has a dual needle with positive headspace pressure, extensive wash routines for minimal carry over and 3 injection modes including µL pick-up mode for zero sample loss. In addition, the micro version of the AS has programmable injection routines called 'User Defined Programs'.

The AS 110 can handle sample volumes down to 1 µL accurately and reproducibly. The sample tray offers temperature control down to 4°C. The fast needle movements result in unsurpassed speeds of less than 30s for a full cycle. These features make the AS 110 a perfect fit for high throughput routine analysis.

The AS110 cool, micro, 10 PV is equipped with an unique stainless steel needle with a total dead volume of 2.4 µl and a micro bore flow path. This allows you to program small sample injections without peak broadening and loss of sample.

 

Part no Description
191.0035 Autosampler AS 110, standard, cool, 6-PV
191.0036 Autosampler AS 110, standard, 6-PV
191.0037 Autosampler AS 110, cool, micro, 6-PV
191.0038 Autosampler AS 110, cool, micro, 10-PV
191.0039 Autosampler AS 110, micro, 6-PV

 

General 

Sound pressure level

LeAq < 70 dB

Working temperature
Storage temperature

10 - 40°C (indoor use only)
–25 - +60°C

Humidity

20 - 80% RH

Safety and EMC compatibility

According to EC-directives, CE, cCSAus (CSA-UL) approved, RoHS compliance

Installation class

II

Pollution degree

2

Altitude

up to 2000 m

Dimensions

300 mm x 510 mm x 360 mm (without cooling option)
300 mm x 575 mm x 360 mm (with cooling option)

Free area around instrument

Minimum free distance of 5 cm from obstacles at rear side and air outlets of the cooling units.

Weight

19 kg (without cooling)
21 kg (with cooling)

Max. weight

that can be placed on top of AS 110: 65 kg

Power requirements

95 - 400 Volt AC ± 10%; 50 - 60 Hz; 200VA

Viscosity range

0.1 - 5 cP

 

Sampling 

Sample capacity

2  Micro Titre Plates according to SBS standards; 96-well high/low and 384-well low formats, 48-vial or 12-vial trays; any combination of plates is allowed, except for 384 Low left and 96 High right.

Vial/Plate dimensions (incl. cap)

Max. plate/vial height: 47 mm (incl. septa or capmat)

Loop volume

Standard: 1 - 5000 µL programmable, 10 mL loop optional
Micro: 0.01 - 100.00 µL programmable with 0.01µL increments

Dispenser syringe

Standard: 500 µL standard or 2500 µL for Prep option
Micro: 25 µL (standard) or 50 µL and 100 µL (optional)

Vial detection

Missing vial/well plate detection by sensor

Headspace pressure

Built-in compressor, but only for vials with septa

Switching time injection valve

Electrically < 100 msec

Piercing precision needle

± 0.6mm

Wash solvent

Integrated wash solvent bottle

Wetted parts in flow path

SS316, PTFE, TEFZEL, VESPEL, Glass, Teflon. Optional: PEEK

Injection cycle time

< 60 sec. in all injection modes for 1 injection ≤ 100 µL including 300 µL wash

 

Analytical performance

Injection modes

Full loop, partial loopfill and µL pickup PASA™ (pressure-assisted sample aspiration)

Reproducibility AS 110
(500 µL syringe)

RSD 0.3% for full loop injections
RSD ≤ 0.5% for partial loopfill injections, injection volumes > 10 µL
RSD ≤ 1.0% for µL  pickup injections, injection volumes > 10 µL
(valid at 1.0 cP)

Reproducibility AS 110 micro (25µL syringe)

 

RSD ≤ 0.3% for full loop injections
RSD ≤ 1.0% for partial loopfill injections, injection volumes > 0.5 µL
RSD ≤ 2.5% for µL  pickup injections, injection volumes > 0.5 µL

Carry-over

< 0.05% with 100 µL needle wash

 

Programming

Interface

AS 110 Service Manager software
DataApex Clarity control module

Injection methods

Full loop, partial loopfill and µL pickup

Injection volume

Standard: 1 µL - 9.999 mL (with 1 µL increment), depending on system settings
Micro: 0.10 µL – 1000.00 µL (with 0.01 µL increment), depending on system settings

Max. injection volume

Full loop  = loop volume
Partial loopfill = ½ x of loop volume
µL Pick up = (loop volume - 3 x needle volume)/2

Injections per vial/well

max. 9 injections

Analysis time

max. 9 hr, 59 min, 59 sec

Wash

Programmable: Wash between injections and Wash between vials

Timed events

Programmable: 4 x AUX ON/OFF

Priority sample

Programmable

 

Communication

Outputs

1 programmable relay output, programmable as Inject marker (default), Auxiliary, Alarm

Inputs

2 programmable TTL inputs, programmable as Next injection input (default), Freeze input, Stop input

Serial communication port

RS232C

 

 

 Options (factory installed)

Sample tray cooling

 

Built-in Peltier cooling
Range: 4°C to Ambient - 3°C
Temp: air temperature in sample compartment: 4°C ± 2°C (at temperature sensor)
(Temperature at relative humidity of 80% and ambient temperature of 25°C)

 

Options (user-installable)

Bio-compatible sample flow path and valve

Inert sample needle (Silco steel) and bio-compatible valve (PEEK)

Prep Kit

2.5 mL syringe, Prep valve, 10 mL sample loop, LSV needle and sample tray for 10 mL vials