AS 110 - Perfect fit for high throughput, |
The AS 110 is a new generation autosampler. It is a fast, robust and easy to use autosampler without compromising accuracy and reproducibility. The AS 110 includes indispensable features for the highest injection performance. It has a dual needle with positive headspace pressure, extensive wash routines for minimal carry over and 3 injection modes including µL pick-up mode for zero sample loss. In addition, the micro version of the AS has programmable injection routines called 'User Defined Programs'.
The AS 110 can handle sample volumes down to 1 µL accurately and reproducibly. The sample tray offers temperature control down to 4°C. The fast needle movements result in unsurpassed speeds of less than 30s for a full cycle. These features make the AS 110 a perfect fit for high throughput routine analysis.
The AS110 cool, micro, 10 PV is equipped with an unique stainless steel needle with a total dead volume of 2.4 µl and a micro bore flow path. This allows you to program small sample injections without peak broadening and loss of sample.
| Part no | Description |
| 191.0035 | Autosampler AS 110, standard, cool, 6-PV |
| 191.0036 | Autosampler AS 110, standard, 6-PV |
| 191.0037 | Autosampler AS 110, cool, micro, 6-PV |
| 191.0038 | Autosampler AS 110, cool, micro, 10-PV |
| 191.0039 | Autosampler AS 110, micro, 6-PV |
|
Sound pressure level |
LeAq < 70 dB |
|
Working temperature |
10 - 40°C (indoor use only) |
|
Humidity |
20 - 80% RH |
|
Safety and EMC compatibility |
According to EC-directives, CE, cCSAus (CSA-UL) approved, RoHS compliance |
|
Installation class |
II |
|
Pollution degree |
2 |
|
Altitude |
up to 2000 m |
|
Dimensions |
300 mm x 510 mm x 360 mm (without cooling option) |
|
Free area around instrument |
Minimum free distance of 5 cm from obstacles at rear side and air outlets of the cooling units. |
|
Weight |
19 kg (without cooling) |
|
Max. weight |
that can be placed on top of AS 110: 65 kg |
|
Power requirements |
95 - 400 Volt AC ± 10%; 50 - 60 Hz; 200VA |
|
Viscosity range |
0.1 - 5 cP |
|
Sample capacity |
2 Micro Titre Plates according to SBS standards; 96-well high/low and 384-well low formats, 48-vial or 12-vial trays; any combination of plates is allowed, except for 384 Low left and 96 High right. |
|
Vial/Plate dimensions (incl. cap) |
Max. plate/vial height: 47 mm (incl. septa or capmat) |
|
Loop volume |
Standard: 1 - 5000 µL programmable, 10 mL loop optional |
|
Dispenser syringe |
Standard: 500 µL standard or 2500 µL for Prep option |
|
Vial detection |
Missing vial/well plate detection by sensor |
|
Headspace pressure |
Built-in compressor, but only for vials with septa |
|
Switching time injection valve |
Electrically < 100 msec |
|
Piercing precision needle |
± 0.6mm |
|
Wash solvent |
Integrated wash solvent bottle |
|
Wetted parts in flow path |
SS316, PTFE, TEFZEL, VESPEL, Glass, Teflon. Optional: PEEK |
|
Injection cycle time |
< 60 sec. in all injection modes for 1 injection ≤ 100 µL including 300 µL wash |
|
Injection modes |
Full loop, partial loopfill and µL pickup PASA™ (pressure-assisted sample aspiration) |
|
Reproducibility AS 110 |
RSD ≤ 0.3% for full loop injections |
|
Reproducibility AS 110 micro (25µL syringe)
|
RSD ≤ 0.3% for full loop injections |
|
Carry-over |
< 0.05% with 100 µL needle wash |
|
Interface |
AS 110 Service Manager software |
|
Injection methods |
Full loop, partial loopfill and µL pickup |
|
Injection volume |
Standard: 1 µL - 9.999 mL (with 1 µL increment), depending on system settings |
|
Max. injection volume |
Full loop = loop volume |
|
Injections per vial/well |
max. 9 injections |
|
Analysis time |
max. 9 hr, 59 min, 59 sec |
|
Wash |
Programmable: Wash between injections and Wash between vials |
|
Timed events |
Programmable: 4 x AUX ON/OFF |
|
Priority sample |
Programmable |
|
Outputs |
1 programmable relay output, programmable as Inject marker (default), Auxiliary, Alarm |
|
Inputs |
2 programmable TTL inputs, programmable as Next injection input (default), Freeze input, Stop input |
|
Serial communication port |
RS232C |
|
Sample tray cooling
|
Built-in Peltier cooling |
|
Bio-compatible sample flow path and valve |
Inert sample needle (Silco steel) and bio-compatible valve (PEEK) |
|
Prep Kit |
2.5 mL syringe, Prep valve, 10 mL sample loop, LSV needle and sample tray for 10 mL vials |