Dialogue Elite

voltammogram

 

 

 

Dialogue Elite for Windows is a multi-functional program to control the DECADE Elite, DECADE II electrochemical detector and the ROXY potentiostat. It runs simultaneously with Windows-compatible integration software or any other Windows program. There are no tools for further data analysis, therefore Dialogue Elite software can not be considered as a fully featured chromatography software package. The software comes as one software package containing functionality for all different versions. Depending on the license dongle extra functionality will be unlocked.

 

Screenshots

Scanning voltammetry
In the scan mode the values can be plotted against potential (voltammogram) or time (chromatogram).

voltammogram

 

PQ noise measurement

 Dialogue can be used for a Performance Qualification using a dummy cell test. The results is a single page PQ report in Excel. Below a typical stabilization pattern. Using the noise monitor enables an automatic restart after 1 min in case the noise is out of spec for convenient unattended operation.

 

noise measurement

 

Timed events

 In Dialogue Elite the operational parameters of the ROXY potentiostat and syringe pump can be controlled through the events table.

Dialogue events table

 

Integrated scripting tool for Calibration

An Calibration script is a series of commands to set parameters and read data from the detector. A script is executed by the script manager in Dialogue software and results are printed through Excel in a calibration report. For successful qualification a calibration kit is available containing cables and tools for testing.

OQ


Microsoft .NET framework

Dialogue Elite requires the Microsoft .NET Framework 4 to be installed. computers connected to internet will automatically install this. Otherwise the full installer version (dotNetFx40_Full_x86_x64.exe) is found on the Microsoft website.

 

Software and drivers for devices are in the download section below.

VSCom usb to serial convertor HASP dongle
USB to serial (RS232) connector VSCom.
The red colored HASP dongle requires a driver installation.



Software

Manuals

A number of support tools are available on this website:

Licences for Dialogue Elite

Part no Description
171.9012 Dialogue Elite OQ, PQ, PM, CAL, ROXY (professional)
171.9015 Dialogue Elite OQ, PQ, ROXY (standard)
171.9017 Dialogue Elite free

The free version is also a trial version which can be downloaded from this web page (download tab). Purchase a license dongle to unlock full functionality if needed, no additional software is required. Contact us for a quote or any other questions. See our website for more information on qualification procedures using Dialogue software.

 

Dialogue free standard professional
Device control + + +
Events + + +
IQ - + +
OQ - + +
PQ - + +
SST - + +
Sequence - + +
Analysis - + +
Scanning - + +
Scripts - - +
Calibration - - +